Meetbereik: | 0.1 - 25.0% O2 |
Nauwkeurigheid: | O2 < 10.0% ± 0.01 bij 2% O2, O2> 10.0% ± 0.1% |
Linearity: | < 0.5% FS |
Vertragingstijd: | < 2 sec |
Reactietijd: | < 5 sec |
Opwarmtijd: | < 30 min |
Nul Drift | ±2% van bereik per maand |
Span Drift | ±2% van bereik per maand |
Voeding: | 100-240V 50-60 Hz |
Energieverbruik: | < 100 VA opwarmen, < 50 VA constant |
Uitgang: | 4-20 mA |
Omgevingstemperatuur: | 0-50ºC |
Max. vermogen analoge uitgang | 500Ω |
Max. vermogen relais contacten | 220 Vac 5 Amp |
Relatieve Vochtigheid: | 90% RH |
Behuizing: | IP 65 stalen behuizing |
Afmetingen (LxBxH) | 380 x 210 x 300 mm met behuizing |
Afmetingen (LxBxH) | 260 x 170 x 220 mm zonder behuizing |
Gewicht: | 10 kg met behuizing. 2 kg zonder behuizing & probe (MARITIME version) |
Certificaat: | DNV |
Max stack gas temperatuur: | 500ºC |
Type sensor: | ZrO2 (zirconia |
Materiaal: | AISI 316 SS |
Bevestiging | 3" withworth pijpdraad DIN ISO 228 |
Lengte: | 250 mm (502 probe)(korte versie) 140 mm (OM10 probe) |
Kabellengte: | max. 10 m (502 probe) max. 2 m (OM10 probe) |
Gewicht: | 2,0 kg (502 probe) (korte versie) 0,2 kg (OM10 probe) |
The Opsis O2000 Oxygen analyser is designed for measuring oxygen in industrial, proces and CEM applications.
Since the oxygen analyser measures the oxygen contents in the flue gas in-situ, there is no need for sample
extraction systems.
The analyser contains an embedded micro controller for fast and easy operation.
It is user friendly with a 2 x 20 characters LCD display with backlight.
One oxygen analyser can handle two O2 probes as an option
Other key features are: Built-in self-checks and diagnostics and built-in reference pump.
- Extended lenght of the probe
- High temperature probe for process temperatures up to 800ºC
- Probe designed for very corrosive environments
- Automatic calibration set O2000 without housing
- Flame arrester to prevent explosion in the process and with OM10 STD Probe
- RS232 output
- Air/Fuel trim function
- High and Low O2 alarm output
- Stainless steel monitor box
- 115 Vac operation
- Stack Unit for cable longer than 10 m
- Alarm Kit 502-probe failure
Model O2000 in cabinet with 502 probe
- Measuring range | 0.1-25.0% O2 |
- Accurracy | O2<10.0% ± 0.01%at 2% O2, O2>10.0% ± 0.1% |
- Linearity | <0.5% FS |
- Lag time | <2 sec. |
- Response time | <5 sec. |
- Warm-up time | <30 min. |
- Zero Drift | ±2% of range per month |
- Span Drift | ±2% of range per month |
- Power supply | 100-240 VAC 50-60 Hz |
- Power consumption | <100 VA warmp up, <50 VA steady state |
- Output | 4-20 mA |
- Max. load, analogue output | 500 Ω |
- Max. load, relay contacts | 220 VAC 5 Amp. |
- Max. ambient temperature | 0-50ºC |
- Max. humidity | 90% RH |
- Enclosure | IP 65 |
- Dimensions old (HxWxD) | 400 x 300 x 210 mm |
- Dimensions new (HxWxD) | 300 x 380 x 210 mm |
- Weight | 10 kg with housing ( 2 kg without housing and probe |
| MARITIME version ) |
- Certification | DNV |
Probe Technical Specifications | |
- Max. stack gas temperature | 500ºC |
- Sensor type | ZrO2 (Zirconia) |
- Material | AISI 316 Stainless steel |
- Mounting | 3" Withworth pipe thread DIN ISO 228 |
- Total lenght | 250 mm (502 probe) (short version) 140 mm (OM10 probe) |
| 730 mm Variable 200-500 mm EPL 1500 max. insert lenght 1500 mm |
- Cable lenght | max. 10 m. (502 probe) excl. stack unit max. 2 m (OM10 probe) |
- Weight | 2,0 kg (502 probe) (short version) 0,2 kg (OM10 probe) |
| 4 kg (502 probe) (long version) |